RECAUTION Is an informative or helpful note. NDICATED The ResMed CPAP S6 system is intended for the treatment of obstructive sleep apnoea (OSA) in adult patients. ONTRAINDICATIONS Nasal CPAP therapy must not be used in patients with the following conditions: •...
Page 7
The clinician should assess on a case-by-case basis the relative risks and benefits of CPAP therapy in such a subject. When assessing the relative risks and benefits, the clinician should understand that the ResMed CPAP S6 unit can be set to deliver pressures up to 20 cmH O.
ResMed mask systems. Other mask systems may be used. However, swings and delivered pressure with non-ResMed masks may be different from those with ResMed masks. If in doubt, measure delivered pressure at the mask. Masks with extremely low exhaust vent flow...
The following are general precautions. Specific precautions appear next to the relevant instructions in the manual. • The patient should never open the ResMed CPAP S6 case. There are no user RECAUTION serviceable parts inside. Repairs and internal servicing should only be performed by an authorised service agent.
CPAP S6 S ERIES Please take some time to become familiar with the features of the ResMed CPAP S6 flow generator. For details on the range of accessories available, see “Accessories/Spare Parts” on page 29. Flow generator Ultra Mirage mask...
CPAP S6 ETTING P THE SSEMBLING THE EVICE Check that you have the following items (see Figure 1, ResMed CPAP S6 system, on page 5): • a ResMed CPAP S6 flow generator unit • a power cord • ResMed air tubing •...
Page 13
ResMed mask systems. Other mask systems may be used. However, swings and delivered pressure with non-ResMed masks may be different from those with ResMed masks. If in doubt, measure delivered pressure at the mask. Masks with extremely low exhaust vent flow...
Page 14
Attach one end of the air tubing firmly to the air outlet on the ResMed CPAP S6 and the other end to the mask system. If the patient is using an axial swivel connector, connect this also. For further information, please refer to the user instructions supplied with the axial swivel.
Page 15
Turn the flow generator switch to ON. The flow generator is now assembled.
A ResControl provides remote setting of Start Pressure, Treatment Pressure, Delay Timers, SmartStart and Humidifier options when it is connected to a ResMed CPAP S6 using the port on the side of the flow generator. ResControl can also display leak and pressure measurements while the flow generator is operating.
Page 17
Foam pad Hour Meter (Plus and Elite only) Carry handle Air outlet The Hour Meter should be reset to zero when the flow generator is given to a patient. You should reset the Hour Meter before setting the Treatment Pressure or Start Pressure.
Page 18
4. Press the Stop button to revert to normal operating mode or turn the flow generator off. ETTING THE REATMENT RESSURE To set the Treatment Pressure, you will need a reference manometer with a range of 0 to 22 cmH O in intervals of 0.1 cmH O, fitted with a sensor tube (2.5 or 3 mm internal diameter) and a standard Luer taper fitting.
Page 19
6. Hold down the Start and 20 minute buttons while you turn on the main power switch at the rear of the flow generator. In the case of the ResMed CPAP S6 Lightweight with a fixed power cord, you will need to plug the power cord back into the power outlet.
Page 20
In the case of the ResMed CPAP S6 Lightweight with a fixed power cord, you will need to unplug the power cord from the power outlet to do this.
Page 21
4. The clock battery has a five year life span. If you are using the flow generator to monitor patient compliance, be sure to contact ResMed to have the battery replaced before the end of the battery life span. If the Real-time Clock is not set, it will still record compliance information but the hours of usage may be allocated to the wrong date when viewing with AutoScan.
Page 22
USING A HUMIDIFIER Set the pressure after connecting the humidifier. If the patient is using a humidifier, then the Humidifier option must be set to ON. 1. Hold down the Start, 5 minute and 20 minute buttons while you turn on the main power switch at the rear of the flow generator.
CPAP S6 SING THE The flow generator should be assembled beside the patient’s bed with the air tubing and mask system connected. (See “Assembling the Device” on page 7.) Have the patient lie down in bed and put the mask on. Arrange the air tubing so that it will move freely if the patient turns in his sleep.
Page 25
TART BUTTON Press the Start button to start the air flow. The air will begin flowing slowly and will build up to full operating pressure in about 10 to 15 seconds. ELAY IMER If patients have difficulty falling asleep with full pressure, they can use the Delay Timer.
In the case of the ResMed CPAP S6 Lightweight with a fixed power cord, you will need to unplug the power cord from the power outlet...
Page 27
6. Hold down the Start, 5 minute, and 10 minute buttons while you turn on the main power switch at the rear of the flow generator. In the case of the ResMed CPAP S6 Lightweight with a fixed power cord, you will need to plug the power cord back into the power outlet.
Always unplug the flow generator before cleaning and be sure that it is dry before plugging it back in. EPLACING THE AIR FILTER STRIPS The ResMed CPAP S6 filter is intended for use with the ResMed CPAP S6 flow generator to filter the air delivered during CPAP treatment. WARNING...
Page 29
4. Reinsert the filter cover into the flow generator. EPLACING THE FILTER ResMed CPAP S6 filters should be replaced every 4 to 6 months, or more often if kept in a dusty environment.
ERVICING All ResMed CPAP S6 series flow generators are designed to give years of trouble- free operation. They will not require regular servicing if they are cleaned and maintained according to the instructions in this manual. If you feel that your unit is not performing properly, follow the directions in “Self- testing (Plus and Elite Models)”...
SmartStart is enabled but the Incompatible humidifier or Use only equipment as flow generator does not stop mask system being used recommended and supplied automatically when you by ResMed remove your mask Flow generator incorrectly Recalibrate (see page 21) calibrated...
Page 32
Problem Possible cause Solution SmartStart not enabled Enable SmartStart (see page 16) Treatment pressure seems low Delay Timer is in use Wait for air pressure to build up gradually, or press the Start button to start immediately Air filter strips are dirty Replace air filter strips Air tubing is kinked or Straighten or replace air tubing...
PECIFICATIONS CPAP S6 F ENERATORS ERFORMANCE Operating pressure range: 4–20 cmH IMENSIONS 110 mm x 240 mm x 290 mm EIGHT 1.77 kg Weight may vary slightly depending on model and country. OWER SUPPLY Input range: 110–240 V, 50–60 Hz, 120 VA...
Page 37
IR TUBING Flexible Hytrel plastic, 2m or 3m length IEC 601-1 C LASSIFICATIONS Class II Type CF The manufacturer reserves the right to change these specifications without notice. GLOSSARY Attention, consult accompanying documents/Attention, consulter les documents d’accompagnement Class II (double insulated)/Classe II (double isolation) Type CF Applied Part/ Partie appliquée du type CF...
Page 38
LOW CHARACTERISTICS RESSURE FLOW FOR ED MASK SYSTEMS Pressure (cm H Mirage Modular, Standard, Disp osable Nasal Mirage Disp Full Face Mirage Full Face Ultra Mirage...
Real-time Clock Start button Self-testing Stopping treatment SmartStart Turning on Humidifier option Treatment Pressure Setting Glossary Troubleshooting Hour Meter Using the ResMed CPAP S6 Resetting Humidifier Setting up Warnings Indicated Use Medical information Plus Adjusting settings using AutoScan Self-testing Precautions...