ResMed CPAP S6 Clinician Manual page 17

Table of Contents

Advertisement

N
OTE
i
N
OTE
i
12
Foam pad
Hour Meter
(Plus and Elite only)
Carry handle
The Hour Meter should be reset to zero when the flow generator is given to a
patient.
You should reset the Hour Meter before setting the Treatment Pressure or Start
Pressure.
To reset the Hour Meter:
1. Hold down the Start, 5 minute and 10 minute buttons while you turn on the
main power switch at the rear of the flow generator. Keep the buttons pressed
until the Start, 5 minute and 10 minute buttons illuminate.
2. Press the Stop and 20 minute buttons together and then release them. The
motor will start running.
At this point, if you press the Start, 5 minute or 10 minute buttons alone, it may
interfere with the flow generator's pressure calibration. If one of these buttons
has been pressed by mistake, all factory defaults will need to be recalibrated.
See "Calibration" on page 21.
3. Press the 20 minute button to reset the Hour Meter.
Air outlet