In-House Factory Calibration; O-Ring Inspection And Replacement; Guidelines For Cleaning Returned Equipment - In-situ Aqua TROLL Care And Maintenance

Advertisement

TROLL Shield Guard (PN 0085190)
The antifouling TROLL Shield Guard is designed to
reduce the fouling of the conductivity cell on Aqua
TROLL 100 and 200 Instruments. Reduced fouling on
the sensor improves accuracy and extends the length of
instrument deployments.
For optimum performance, the TROLL Shield Guard
should be replaced every 6 months.
TROLL Shield Guard

O-ring Inspection and Replacement

Examine O-rings for wear, dryness, discoloration,
stretching, cracks, nicks, and brittleness. Replace when
any of these conditions are present. Replacing O-rings
on a regular schedule, regardless of O-ring condition, is
the best way to protect against damage.
Perform the following steps to replace an O-ring.
1. Remove and discard the damaged O-ring.
2. Use a clean, dry soft cloth to clean the O-ring
groove to remove dirt or residue.
3. Lubricate the new O-ring using high-vacuum
grease.
a. Wash your hands thoroughly.
b. Apply a small amount of grease to the pad of
your index finger, and rub your index finger
and thumb together to spread the grease
evenly.
c. Inspect the new O-ring and remove any
debris stuck to it.
d. Rub your fingers around the O-ring until there
is a thin layer of grease on the entire O-ring
body.
4. Install the O-ring in the groove and remove any
excess lubricant with a clean cloth.
Do not allow water or lubricant to enter the
connector.
800-446-7488
Factory Calibration Services

In-House Factory Calibration

Factory calibration of In-Situ Inc.instruments should be
Instruction Sheet
performed every 12 months, or when the unit data
appears to drift significantly. Factory calibration includes
a thorough cleaning, full functionality check, and
adjustments of the pressure sensor, temperature
sensor, and conductivity sensor (when applicable) over
the entire calibrated temperature range.
To obtain a factory calibration, fill out and return the
online RMA (Return Materials Authorization) form
located at www.in-situ.com/Return_authorization.

Guidelines for Cleaning Returned Equipment

Please help us protect the health and safety of our
employees by cleaning and decontaminating equipment
that has been subjected to any potential biological or
health hazards, and labeling such equipment.
Unfortunately, we cannot service your equipment
without such notification. Please complete and sign the
form in your Operator's Manual (or a similar statement
certifying that the equipment has been cleaned and
decontaminated) and send it with each returned
instrument.
We recommend a cleaning solution, such as
l
®
Alconox
, which is a glassware cleaning product
available from In-Situ Inc. (PN 0029810) or
laboratory supply houses.
Clean all cabling. Remove all foreign matter.
l
Clean cable connector(s) with a clean, dry cloth.
l
Do not submerge cable connectors.
Clean the probe body—including the nose cone,
l
cable head, and protective caps. Remove all
foreign matter.
If an instrument is returned to our Service Center for
calibration or repair without a statement that it has been
cleaned and decontaminated, or in the opinion of our
Service Representatives presents a potential health or
biological hazard, we reserve the right to withhold
service until proper certification has been obtained.
5