Cl 100 S Urgical Illuminator Disassembly; Power Supply Removal - Welch Allyn CL 300 Service Manual

Surgical illuminator
Table of Contents

Advertisement

CL 100 S
I
URGICAL
LLUMINATOR
P
S
R
OWER
UPPLY
EMOVAL
1. Remove the 9 screws securing the top cover and remove the cover.
2. Remove the 3 screws from the bottom side of the back panel.
3. Disconnect wires from J1 and J2 on the power supply. Disconnect fan connector from J4.
4. Remove the 2 screws from the bottom chassis, securing the center baffle (see figure below).
Screws
securing center
baffle to bottom
chassis
5. Remove the nut that secures the center baffle to the back panel.
6. Pull back panel out and move to the platform side of the chassis to gain access to the back of the power
supply.
PN: LB-MAN-CLSERV Rev. A
D
ISASSEMBLY
1-15