Welch Allyn CL 300 Service Manual page 50

Surgical illuminator
Table of Contents

Advertisement

2. After the fifteen (15) minute burn-in insert the POWER TEST FIBER into the Port. Open the attenuator fully
and measure the light output. Rotate the proximal end to maximize power output and check the following:
Does the connector insert smoothly?
Does the attenuator move smoothly?
Does the Light Output measure > 361 lumens? Light output < 200 is considered the end of the life of the
lamp.
1-44