Service; Cleaning; Surgical Illuminator; Multi Link ™ Expansion Port - Welch Allyn CL 300 Service Manual

Surgical illuminator
Table of Contents

Advertisement

S
ERVICE
C
LEANING
1. Prior to cleaning, turn the power switch OFF, and disconnect power source from both the Light Source and
the wall outlet.
2. Clean the external surfaces by wiping with a cloth dampened with 10% bleach solution. DO NOT IMMERSE.
Routinely inspect and clean all air intakes / out flows at the rear of the chassis for lint or other debris.
Follow the manufacturer's instructions when Cleaning the fiber bundles or accessories used with the light
source.
3. Wipe the power cord with a cloth dampened with 10% bleach solution and water. DO NOT IMMERSE. DO
NOT RECONNECT WET.
4. DO NOT wipe down any lenses or windows.
5. DO NOT plug the power source into a wall outlet until it is thoroughly dry.
S
I
URGICAL
LLUMINATOR
The console may be cleaned using a clean soft cloth with the following cleaning agents:
Enzymatic Solution (Enzol)
Isopropyl Alcohol
Hydrogen Peroxide 3% USP
2% Glutaraldehyde Solution
Soap and Water
10% bleach solution and water
CAUTION: Do not splash cleaning agents into any openings, seams or electrical components.
M
L
™ E
ULTI
INK
XPANSION
The outside surfaces may be cleaned with a soft cloth using the following:
Isopropyl Alcohol
Soap and Water
The glass surfaces may be cleaned by the following methods:
A cotton swab moistened with Isopropyl Alcohol
Blow with dry compressed air
NOTE: Use care in cleaning the optical surfaces of the Multi Link™ Expansion Port to prevent scratches.
PN: LB-MAN-CLSERV Rev. A
P
ORT
1-9