High Level Calibration - MAC3 Sensopress LCD User Manual

Table of Contents

Advertisement

Sensopress_LCD_Rev_00_Eng.doc Eng
Figure 16
By pushing the Yes/+ button it is possible to make the zero setting and on the display it is shown the
writing SAVING.......for a few seconds, after which the device turns in the AUTO Mode.
By pushing the Menu button there is the possibility to go out from the calibration procedure without
saving the zero. It turns back to the AUTO mode.
5.1.2

High Level Calibration

It is possible to enter the calibration of high level by pushing the Yes/+ button after entered the CALIB.
Make sure the sensor is put through the water column which needs to be set-up.
On the display is visualized the screen of Figure 17.
Figure 17
With the Yes/+ and No/- buttons it is possible to set-up the water column value to which the high level
calibration needs to be done and to which is put through the sensor.
Push the Menu button to enter the saving of the high setting level and it is visualized the screen of
Figure 18.
Figure 18
By pushing the Yes/+ button it is possible to do the saving of the set-up high level and on the displays
it is given the writing SAVING..... for a few seconds, after this the device turns in the AUTO mode.
18