Setting/Changing The Datarecorder Mode; ("Off", "Manual", "Auto") - Dräger CMS Instructions For Use Manual

Analyzer with integrated datarecorder
Hide thumbs Also See for CMS:
Table of Contents

Advertisement

Available languages

Available languages

Preparing for use
6.4

Setting/changing the DataRecorder mode

("Off", "Manual", "Auto")

The measured values are saved by the DataRecorder integrated into the analyzer of the Dräger
chip measurement system. The following storage modes can be selected:
"Off" - Switch memory function off.
"Man" - Results can be saved by hand with assignment of a location code.
"Auto" - Automatic storage - without confirmation and without assignment of a location
code.
Results can be saved with time, date and location code (only in "Man" mode). This data can
then be recalled and displayed.
50 results can be saved. The data in the first memory location will be overwritten if more than
50 results are to be saved.
A corresponding warning is displayed in this case (with the 51st result).
1. Move sliding switch from position 0 to
position 1. Then immediately press and
hold side button until the following display
appears:
Setup CMS
Read DataRec
2. Briefly press side button several times
until the following display appears:
Setup CMS
DataRec Mode
3. Press and hold side button until the
following display appears:
DataRec Mode
off / man / auto
The effective storage mode is underlined.
Set storage mode:
1. Briefly press side button several times until the required storage mode is underlined.
Dräger CMS Analyzer
Setup CMS
Read Data Rec
0
0
1
1
2
2
3
3
Setup CMS
Data Rec Modus
DataRec Mode
off / man / auto
10022837_4en.eps
01622837.eps
01722837_4en.eps
43

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents