Configuring The Communications Protocol - Hioki SM7860-01 Instruction Manual

Table of Contents

Advertisement

3.4 Configuring the Communications Protocol

Configuring GP-IB Interface Communications
Setting the address
1
Press and hold the scroll keys (
for about 7 seconds. (The address can be set from the P1 or P2 screen.)
Screen P1: Status display
V A ( + ) I R : 1 0 0 0 . 0
V B ( + ) I R :
T E R M I N A L :
O U T 1 : 1 1 1 0 0 0 0 0
O U T 2 : 1 1 1 0 0 0 0 0
O U T 3 : 1 1 1 0 0 0 0 0
O U T 4 : 1 1 1 0 0 0 0 0
2
Using the scroll keys (
(Valid setting range: 1 to 30)
G P I B A D D R : 1
3
When finished making the setting, turn off the device.
4
Turn on the device.
The device will revert to the initial screen, and the GP-IB address will be set to the
selected address.
Configuring RS-232C Interface Communications
Communication conditions
Remote switching requests
PAGE UP▲
2 5 0 . 0
V M A : 1 0 0 0 . 0
V M B :
/
PAGE UP▲
PAGE DOWN▼
Baud rate
Parity
Stop bit
Data
Flow control
Send the "
RMT
" command from the RS-232C interface.
RS-232C interface communications will not be available until the "
mand is sent.
3.4 Configuring the Communications Protocol
/
) on the front of the device
PAGE DOWN▼
O K
2 5 0 . 0
O K
), set the desired address.
38400 bps
none
1 bit
8 bits
none
21
RMT
" com-

Advertisement

Table of Contents
loading