18

3.2 Specifications

3.2 Specifications
GP-IB Specifica-
tions
RS-232C Specifica-
tions
Precautions
RS-232C and GP-IB communications cannot be used simultaneously.
Electrical machinery specifications: IEEE std. 488.1-1987 compliant
Address setting
: Can be set to talker/listener addresses 1 to 30.
Interface Functions
SH1
All Source Handshake functions
AH1
All Acceptor Handshake functions
T6
Basic talker functions
Serial poll function
Talk-only mode
The talker cancel function with MLA (My Listen Address)
L4
Basic listener functions
Listen-only mode
The listener cancel function with MTA (My Talk Address)
SR1
All Service Request functions
RL1
All Remote/Local functions
PP0
Parallel Poll function
DC1
All Device Clear functions
DT1
All Device Trigger functions
C0
Controller functions
E2
Tri-state output
Operating Code: ASCII codes
Transfer method
Baud rate
Data length
Parity
Stop bit
Flow control
Electrical specification
Connector
Operating Code: ASCII codes
Communications: Full duplex
Synchronization: Start-stop synchronization
38400 bps
8 bits
none
1 bit
none
Input voltage levels
5 to 15 V
-15 to -5 V
Output voltage levels
+5 V or more : ON
-5 V or less
RS-232C Interface Connector Pinout
(Male 9-pin D-sub, with #4-40 attachment screws)
The I/O connector is a DTE (Data Terminal Equipment)
configuration
Recommended cables:
Model 9637 RS-232C Cable
Model 9638 RS-232C Cable
See:
"3.3 Connect a cable to the GP-IB connector or
RS-232C connector" (p.19)