2.3 Connecting the Device to the Measuring
Instrument
Connection Methods
1
Confirm that the device is turned off.
2
Connect the included voltage output con-
nector to the voltage output terminal on
the rear of the device.
Specifications
Voltage output pin assignments
2
3
1
7
6
5
4
8
10
9
(View of terminal on device)

2.3 Connecting the Device to the Measuring Instrument

To avoid electric shock or damage to the equipment, always observe the
following precautions when connecting to voltage output terminal.
• Always turn off the power to the device and to any devices to be con-
nected before connecting the voltage output connectors.
To avoid damaging the device, be sure to observe the following precautions:
• Do not connect the device to a load with a charge in excess of the output volt-
age range.
• Do not connect the device to a load with a charge that has the opposite polarity
of the output voltage range.
• Do not connect the device to a load that has a charge when the device's power
supply is turned off.
Pin No.
Circuit A
1
2
3
4
5
6
7
8
9
10
*1 On the SM7860-05, -06, -25, and -26, serves as the discharging
terminal.
*2 On the SM7860-05, -06, -07, -25, -26 and -27, serves as the dis-
charging terminal.
Voltage output terminal
*1
OUT1
OUT2
Circuit A
CH1
CH1
CH2
CH2
CH3
CH3
CH4
CH4
CH5
CH5
CH6
CH6
CH7
CH7
CH8
CH8
NC
NC
COM
COM
Rear Panel
*2
OUT3
OUT4
Circuit B
Circuit B
CH1
CH1
CH2
CH2
CH3
CH3
CH4
CH4
CH5
CH5
CH6
CH6
CH7
CH7
CH8
CH8
NC
NC
COM
COM
13