7.4.4

BPM module error messages

Error message
BPM: Measurement value
outside the alarm limits
BPM: Measurement not
successful
BPM: Option not available
E02 BPM error
(Remove the cuff from the
patient!)
E03 BPM error
(Remove the cuff from the
patient!)
E06 BPM error
(Remove the cuff from the
patient)
BPM: Cuff cannot be inflated.
BPM: Inflation time limit
exceeded
BPM: Cuff pressure lower
than preselected pressure
BPM: Measuring time limit
exceeded
BPM: Maximum cuff pressure
exceeded
BPM: Inflation pressure
reached too quickly.
Exx BPM error
(Remove the cuff from the
patient!)
E25 BPM error
(Remove the cuff from the
patient!)
E26 BPM error
(Remove the cuff from the
patient!)
E70 BPM error
(Remove the cuff from the
patient!)
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
Description
Blood pressure cuff not correctly connected. Leakage in pressure tubing. Alarm
limits not adjusted to the patient.
It was not possible to determine a measurement value.
BPM option not installed.
Due to the error FXX BPM error the BPM option is no longer available during
the current treatment.
Module does not respond anymore
Module reset
BPM Rcv buffer vent
CPU-P1 pressure < 25 mmHg 15 seconds after the start of a measurement
Inflation time > 70 seconds
CPU-P1 pressure < 40 % of the desired pressure
Measurement time > 175 seconds
CPU-P1 pressure > 320 mmHg
Inflation pressure reached too quickly.
Sample array overflow
PulsCount array overflow
StepCount array overflow
Auto-zero error.
A failure is detected if the CPU-P1 pressure is > ±9 mmHg or the pressure
change is > ±1 mmHg 5 seconds after starting a measurement.
Chapter 7: Error messages
7-35