Module Options Screen; Module Options - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Appendix E

Module Options Screen

Figure 118 – Module Options Screen (showing functional software version 2.70 and
BlueStar Premium)
PatientCard
Premium)
This button enables or disables the optional PatientCard hardware. See page 37 for more
information.
Note: The 'PatientCard' option is part of BlueStar Premium and requires additional hardware.
Contact the Fresenius Medical Care Spare Parts department or your sales representative for
more information and pricing.
Crit-Line
The Crit-Line option must be set to 'Yes' in Service Mode in order to use the Crit-Line in a
Clip (CLiC) device during a hemodialysis treatment. Note: The BTM and BVM module
options are unavailable if Crit-Line is selected.
322
(functional software version 2.70 or later with BlueStar
(functional software version 2.57 or later)
2008T Machine Operator's Manual P/N 490122 Rev Q
TM

Advertisement

Table of Contents
loading

Table of Contents