5.2.2
P.C.B. LP 1631 (CPU 1) DIP switch array 2
DIP switch /
Position
1
ON
OFF
2
3
ON
ON
OFF
ON
ON
OFF
OFF
OFF
4
ON
OFF
5
ON
OFF
6
ON
OFF
7 *
ON
OFF
8
ON
OFF
The basic position on delivery is in italics.
"No assignment" requires the OFF position.
* DIP switch 7 is no longer relevant if DIP switch 8 is set to ON.
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
S1
Function
CAL mode
Mode 0
Mode 1
External alarm input
Not valid
RO system
Patient bell
External alarm
Remote control
Device with remote control
Device without remote control
COMMCO P.C.B. LP 763
Approved
Disabled
COMMCO
Special protocol
Default protocol
Rinse section test (exclusive CDS)
Active
Not active
Central Delivery System
Installed
Not installed
S2
S3
Chapter 5: Setup / service program
LP1631
5-65