Troubleshooting - Thermo Scientific 05EFEETSA Installation And Operation Manual

General purpose, flammable material storage and explosion proof freezers
Hide thumbs Also See for 05EFEETSA:
Table of Contents

Advertisement

In the event that your unit is not operating properly, check the following before calling for service assistance this
may save you the cost of unnecessary service calls.
Symptom
Unit not operating.
Unit runs continuously.
Noise problems.
20
Possible Cause of Problem
Make certain that the unit is connected to a grounded outlet.
Make certain that temperature control knob is ON.
Check that circuit breaker is not tripped or fuse is not blown.
Make certain that there is no heavy frost accumulation. If there is, defrost
unit.
Look at condenser to see if there is layer of dust or lint. Clean if required
with a dry brush or vacuum.
A leak around the door gasket will allow cold air to escape. This causes unit
to work harder than necessary to maintain cold temperatures. Re-seat or
replace the gasket if worn.
If you have temperature set too cold, this may cause unit to run
continuously. Check optimum running temperature.
Is the ambient air over 43ºC (109ºF), or the units located close to
heat sources? If possible, move to a different location.
An unusually high frequency of door openings and closings can increase
operating load. Unit will stabilize as these are decreased.
This can be caused by contents of unit being set too close and rattling
against each other. Rearrange contents as needed.
Hissing or gurgling noise is caused by refrigerating fluid circulating and is
normal.
Noise can result if unit is not level on floor. Check with level.
Fan noise: normal airflow can cause this, not a problem.

Troubleshooting