Operation; Environmental Operating Conditions; Start-Up Procedure - Thermo Scientific 05EFEETSA Installation And Operation Manual

General purpose, flammable material storage and explosion proof freezers
Hide thumbs Also See for 05EFEETSA:
Table of Contents

Advertisement

Operation

Environmental Operating Conditions

POLLUTION DEGREE*:
INSTALLATION CATEGORY*:
ALTITUDE:
HUMIDITY:
ELECTRICAL SUPPLY:
VOLTAGE TOLERANCE: AMBIENT
TEMPERATURE RANGE:
PRODUCT USAGE:
* Refer to IEC 664-1
18
2
II
2000 Meters MSL (Mean Sea Level)
80% maximum, non-condensing
120VAC or 240VAC
±10% of normal rated line
15ºC to 40ºC
This product is intended for use indoors only

Start-Up Procedure

The markings on the thermostat knob range from 1
through 6, with 1 being the warmest setting and 6
the coldest.
When the unit is turned on for the first time (or after
defrosting), set the control to its maximum position,
6, and leave it for at least two hours before
introducing the material to be stored.
After 24 to 48 hours, adjust the temperature control to
the setting that best suits your needs. A setting of 3~4
would be recommended.
NOTE: If the unit is unplugged, has lost power, or is
turned off, you must wait 3 to 5 minutes before
restarting the unit. If you attempt to restart before this
time delay, the freezer will not start.