Clean The Gas-Ballast Filter; Figure 9 Remove/Replace The Inlet Filter - Edwards E2M28 Instruction Manual

Rotary vacuum pumps
Hide thumbs Also See for E2M28:
Table of Contents

Advertisement

Figure 9 Remove/replace the inlet filter

1. Inlet adaptor
1. Inlet adaptor
3. Circlip
3. Circlip

6.6 Clean the gas-ballast filter

Note:
The gas-ballast filter may have been removed to connect a gas supply or valve to the
gas-ballast inlet.
Refer to
1.
Remove the retainer circlip.
2.
Remove the protective wire mesh and filter elements.
3.
Wash the wire mesh and filter in a suitable cleaning solution. Allow the wire mesh
and filter to dry before replacing them in the pump.
4.
Refit the wire mesh and filter element in the adaptor plate housing and retain with
the retainer circlip.
A37310880_V - Maintenance
Figure: Remove/replace the gas-ballast filter on page
1
2
3
4
2. O-ring
2. O-ring
4. Inlet-filter
4. Inlet-filter
Page 36
37.