HistoStar Operator Guide A81010100 Issue 2

General Instrument Cleaning

To ensure that this instrument remains in peak condition, it is strongly recommended
that it is serviced annually by a Thermo Fisher Scientific approved Service Engineer.
When cleaning the surfaces of HistoStar , ensure any necessary precautions are taken
to avoid cross-contamination of specimens.
Xylene has a flash point of 27°C - 32°C and is therefore not suitable for use when
cleaning the HistoStar .
Cleaning the Casework and Cold Plate
To clean the casework and cold plates:
Use the plastic scraper supplied to remove hardened wax from the surfaces.
Once the wax has been removed, use a damp, lint-free cloth to wipe the surfaces.
Cleaning the Heated Surfaces
To clean the heated surfaces:
Use a dry, lint-free cloth to wipe the surfaces clean.
Pay particular attention to the drainage channels.
Clean the forceps block holes to ensure that any debris is removed.
Note
A cotton tipped swab is useful for cleaning the drainage channels.
Avoid skin contact with the heated surfaces as painful injuries can occur.
Cleaning the Touchscreen
To clean the touchscreen:
Use a dry, lint-free cloth to wipe the touchscreen clean of fingerprints.
If any wax has dried onto the touchscreen surface, carefully remove it using the plastic scraper supplied.
Do not use mineral acids, abrasives, benzyl alcohol or high pressure steam to
clean the touchscreen.
92