Agilent Technologies VS C15 Series Field Replacement Instructions page 8

Calibrated leak replacement for leak detectors

Advertisement

11. Reattach the leak assembly to the flange with the clamp.
12. Close the cover and secure using the existing hardware.
13. Power up the VS C15 component leak detector.
14. Set up the Internal Calibrated Leak by either:
Setting up the Internal Calibrated Leak Option via the front panel display by:
a. Select Advanced Parameters > Internal Type and select STD LEAK.
b. Select Maintenance > Internal Calibrated Leak and input the calibration data from the Calibration Certificate
for the internal calibrated leak including:
• Internal Leak Value
• Temperature
• Temperature coefficient factor
• Cal Leak - Date of Expiration
c. Validate the setup using the Calibration Set Up menu: the temperature compensated leak value, internal
cali-brated leak temperature and date of expiration.
— or —
• Setting up the Internal Calibrated Leak Option via the RS232 serial communications port (see the operation
manual for communicating via the RS232 serial port) by:
a. Enter the following commands:
• 1 INIT-INTERANL-TYPE
• X.XE-0X INIT-STDLEAK (Enter the leak rate value from the calibration certificate ex. 1.8E-07 INIT-STDLEAK).
• XX.X INIT-LEAKTEMP (Enter the temperature from the calibration certificate ex. 23.5 INIT-LEAKTEMP).
• -/+XX INIT-TEMPFACTOR (Enter the temperature coefficient from the calibration certificate ex. -7 INIT-
TEMPFACTOR).
• mm dd yyyy INIT-LKEXPIRE (Enter the date of expiration from the calibration certificate ex. 12 22 2008 INIT-
LKEXPIRE).
b. Validating that the correct internal leak data was inputted (data from the Calibration Certificate) by:
• ?STDLEAK (reports back the value of the internal calibrated leak).
• ?LEAKTEMP (reports back the temperature of the calibrated leak from the factory calibration).
• ?TEMPFACTOR (reports back the temperature coefficient).
• ?LKEXPIRE (reports back the date of expiration).
Perform an internal calibration (via the I/O, Front Panel Display or RS232) to validate a successful installation of
the internal calibrated leak.
Calibrated Leak Replacement for VS C15, VS, and (G8610, G8611, or G8612) Series Leak Detectors
Page 6 of 12