2.3 Control Box

The control box has been specially developed for patient lifting devices and the
complete system is medically approved and contains a series of features that meet
the patients needs for a safe and comfortable transfer e.g. soft start function for
raising and lowering.
The control box has a built in charging system and so removes the issues of lost or
damaged chargers. To recharge the batteries the hoist is connected to the mains
via a mains lead that is made from a curly flex to give a greater length of cable
without the need for trailing cables on the floor. See section 'Re-charging the
batteries'
The actuators for the raising / lowering and the leg spreading are connected using
jack plugs into the bottom of the control box..
Power 'ON'
Indicator
Handset
Connection
L1 - Actuator Connection
LCD Display
The LCD display has a range of options available and can be viewed during the
operation of the hoist.
By pressing the "UP" button on the handset the LCD display indicates the following:
(a)
How many lifts the hoist has completed
(b)
How many times the lift has been overloaded
(c)
Service interval.
When the hoist is being used the LCD will also display the function ie, raising,
lowering, opening &/or closing the legs and battery charge level. See photographs
on page 8.
See page 11 for recharging the batteries.
SA160 User Manual
L2 - Leg Actuator Connection
Rev 5—October 2016
Charging Indicator
LCD Display
Mains
Connection
Page 7