Adjust:clear; Adjust; Sample:rate <Slow|Medium|Fast|Exfast - ITech IT5101 Programming Manual

Internal resistance tester
Hide thumbs Also See for IT5101:
Table of Contents

Advertisement

Arguments
1|0|ON|OFF

ADJust:CLEAr

This command is used to clear the state of adjusting to zero.
Command syntax
ADJust:CLEAr
Query syntax

ADJust?

Arguments
NONE
ADJust?
This command is used to adjust to zero and query the result.
Command syntax
ADJust?
Returns
0|1
0:OK
1:Failure
Note
SAMPle:RATE <SLOW|MEDium|FAST|EXFast>
This command is used to set or query the sample speed.
Command syntax
SAMPle:RATE <SLOW|MEDium|FAST|EXFast>
Query syntax
SAMPle:RATE?
Arguments
SLOW|MEDium|FAST|EXFast
Example
SAMPle:RATE SLOW
The instrument needs a while for adjusting to zero. Please check the returns after setting
the time interval or set the overtime interval to 8 seconds..
Copyright© Itech Electronic
Configuration Commands
Co., Ltd.
22