Chapter 18
NIBP SPECIFICATION
Name
Display mode
The degree of protection against ingress of water
NIBP specification
Measurement method
Working mode
Operation mode
Measurement Range
Resolution
Accuracy
Error
Operating temperature/ humidity
Transport
Storage
ENGLISH
Electronic Sphygmomanometer
2.8'' color LCD Display
IPX0
Oscillometric method
Automatic
Continuous operation
Pressure:0~290 mmHg (0~38.6 kPa)
Pulse: 40~240/min
Pressure: 1mmHg(0.133 kPa)
Static pressure: ±3 mmHg (±0.4 kPa)
The BP Value of the device is equivalence with the measurement
value of Stethoscopy. The error meets the requests in the ANSI/
AAMI SP-10:2002+A1:2003 +A2:2006
+5 ºC~40 ºC
Transport by general vehicle or according to the order contract,
avoid pounded, shake and splash by rain and snow in transpor-
tation.
Temperature: -20 ºC~+55 ºC;
Relative humidity: ≤95%; No corrosive gas and drafty.
15%RH~85%RH (Non-condensing)
74

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CONTEC08A-VET and is the answer not in the manual?

Questions and answers

Lisa
March 20, 2025

When I take the BP on my dog, it just keeps going and never gives a pressure. Cuff is on properly, is the right size and air plug is in securely. What **** I doing or not doing correctly?

1 comments:
Mr. Anderson
March 20, 2025

Several factors could prevent the CONTEC08A-VET from providing a blood pressure reading on your dog despite proper cuff placement and secure connections:

1. Animal Condition: The device should not be used on animals with damaged skin, sickle-cell disease, or thrombasthemia without clinical evaluation.
2. Incorrect Cuff Size or Pressure: Ensure that the correct cuff size is used and that it is not overinflated, as this may prevent accurate readings.
3. Movement or Stress: If the animal is moving or stressed, it can interfere with measurements. Keep the animal calm and still.
4. Device Lifespan: If the device has exceeded its five-year useful life, it may no longer function properly.
5. Environmental Factors: Ensure the device is not used in conditions with flammable anesthetic gases or if liquid has entered the device.
6. Software or Connection Issues: If using the USB interface to transfer data, check the related software for proper operation.

If issues persist, consult the manual for troubleshooting or contact the service department.

This answer is automatically generated