Chapter 15 Troubleshooting - CONTEC MEDICAL SYSTEMS CONTEC08A-VET User Manual

Vet electronic sphygmomanometer
Table of Contents

Advertisement

69
Chapter 15
TROUBLESHOOTING
Abnormal Phenomenons
BP measurement values too high
or too low
No pressure
Cuff deflates in short time
It can not carry on measurement when press the measurement button
Power off suddenly when inflating
Hold the on/off button but can not
start the device
Cuff inflation start before press the measurement button or never stop
inflating when measuring
Cuff never deflation
Causes
Cuff is not connected correctly
Move limbs when measuring
Cuff leakage
The cuff windpipe is not correctly
connected with cuff
Cuff is not inflated
Loose cuff
No use for a long time, the power of
batteries can be exhausted owing to
the changed temperature
Power of batteries can be exhausted
The battery polarities is reversed
ENGLISH
Solutions
Correctly connect cuff (refer to Chapter 9)
Keep quiet and restart a measurement
Buy a new cuff
Correctly connect
Stop using the device and contact us
Correctly apply cuff
Switch on the power once again and
restart a measurement
Replace all four batteries with new ones.
Replace all four batteries with new ones
Check the battery installation for proper
placement of the battery polarities.
Pull out the cuff to deflate. Stop using the
device and contact us
Pull out the cuff to deflate. Stop using the
device and contact us

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CONTEC08A-VET and is the answer not in the manual?

Questions and answers

Lisa
March 20, 2025

When I take the BP on my dog, it just keeps going and never gives a pressure. Cuff is on properly, is the right size and air plug is in securely. What **** I doing or not doing correctly?

1 comments:
Mr. Anderson
March 20, 2025

Several factors could prevent the CONTEC08A-VET from providing a blood pressure reading on your dog despite proper cuff placement and secure connections:

1. Animal Condition: The device should not be used on animals with damaged skin, sickle-cell disease, or thrombasthemia without clinical evaluation.
2. Incorrect Cuff Size or Pressure: Ensure that the correct cuff size is used and that it is not overinflated, as this may prevent accurate readings.
3. Movement or Stress: If the animal is moving or stressed, it can interfere with measurements. Keep the animal calm and still.
4. Device Lifespan: If the device has exceeded its five-year useful life, it may no longer function properly.
5. Environmental Factors: Ensure the device is not used in conditions with flammable anesthetic gases or if liquid has entered the device.
6. Software or Connection Issues: If using the USB interface to transfer data, check the related software for proper operation.

If issues persist, consult the manual for troubleshooting or contact the service department.

This answer is automatically generated