Chapter 14 Error Message - CONTEC MEDICAL SYSTEMS CONTEC08A-VET User Manual

Vet electronic sphygmomanometer
Table of Contents

Advertisement

Open Windows Explorer.
Click on the root CD-ROM directory.
Double click file setup software.
Follow the instructions in the screen.
Refer to "Software Help"for details about the operation method of the PC software.
Chapter 14
ERROR MESSAGE
Error message will be displayed in the screen if there is something wrong when measuring. The causes and solutions are
shown as follows:
Error Message
Self-test failure
System failure
Loose cuff
Air leakage
Air pressure error
Weak signal
Overpressure
Excessive movement
Over range
Saturated signal
Time out
ENGLISH
Causes
Function abnormal
Cuff is not connected correctly.
Cuff plug fall off
Air pressure error
The pulse signal is too weak or the cuff
is loose.
Cuff is blocked or squeezed
The signal extent is too big owing to the
limbs or body moving or other reasons
when measuring
It takes too much time
Solutions
Please contact us
Correctly connect cuff (refer to Chapter 9)
Make sure the cuff plug is securely inserted
in the windpipe (refer to Chapter 9)
Refer to the troubleshooting
Correctly connect cuff (refer to Chapter 9)
Correctly connect cuff (refer to Chapter 9)
Keep limbs, body still, measure again
68

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CONTEC08A-VET and is the answer not in the manual?

Questions and answers

Lisa
March 20, 2025

When I take the BP on my dog, it just keeps going and never gives a pressure. Cuff is on properly, is the right size and air plug is in securely. What **** I doing or not doing correctly?

1 comments:
Mr. Anderson
March 20, 2025

Several factors could prevent the CONTEC08A-VET from providing a blood pressure reading on your dog despite proper cuff placement and secure connections:

1. Animal Condition: The device should not be used on animals with damaged skin, sickle-cell disease, or thrombasthemia without clinical evaluation.
2. Incorrect Cuff Size or Pressure: Ensure that the correct cuff size is used and that it is not overinflated, as this may prevent accurate readings.
3. Movement or Stress: If the animal is moving or stressed, it can interfere with measurements. Keep the animal calm and still.
4. Device Lifespan: If the device has exceeded its five-year useful life, it may no longer function properly.
5. Environmental Factors: Ensure the device is not used in conditions with flammable anesthetic gases or if liquid has entered the device.
6. Software or Connection Issues: If using the USB interface to transfer data, check the related software for proper operation.

If issues persist, consult the manual for troubleshooting or contact the service department.

This answer is automatically generated