CONTEC MEDICAL SYSTEMS CO.,LTD
Product name:
Pulse Oximeter Probe
Product model:
ESA0008,ESA0015,ESC0029,ESA0004,ESA0014,ESA0005,ESA0016,ESA0061
Scope of application:
This product is used to match CONTEC Patient Monitor 、Pulse Oximeter and Electronic sphygmomanometer ,collect and transmit the
SpO
signal from patient with continuance and no trauma. It is inapplicable to monitor the weak perfusion moving state and monitor for
2
long, so check the measuring position or change for another position per 4 hours.
Taboo disease:
Don't fix the product on the position with tissue injury. It is inapplicable for the patient or users allergic to PVC、TPU、TPE、ABS plastic.
Product performance:
1) The range of SpO
measurement: 70%~100%;
2
Accuracy: 70~100%:±2%; Below 70%: unspecified.
2) The range of pulse measurement: 30~250bpm;
Accuracy: ±2bpm or ±2%(select larger).
3) Optical Sensor:
Red light (wavelength is 650~670nm, 6.65mW)
Infrared (wavelength is 880~910nm, 6.75mW)
Main configuration:
Consisting of plug, cable and probe.
Power supply requirement:
and Electronic sphygmomanometer which are applicable to the requirements of IEC60601-1.
Directions for use:
Note:
This product is type BF applied part;
Sketch map
1) As Figure 1, the pulse oximeter probe of different types is applied to different crowds.
2) Select proper probe and put recommendatory placement according to Figure 1.
3) Arrange the cable along the back of hand when place the pulse oximeter probe.
4) Connect Pulse oximeter probe with Pulse oximeter ,Patient Monitor or Electronic sphygmomanometer and check if the operating
procedure accords with the procedure introduced in user manual.
5) Pulse Oximeter probe ESC0029 needs the help of the FST0001 Pulse Oximeter probe extension cable to be connected in to the jack
of the Pulse Oximeter CMS60D,CMS70A or Patient Monitor CMS8000(old model),Pulse Oximeter probe ESC0029 needs the help of the
FST0004 Pulse Oximeter probe extension cable to be connected in to the jack of the Electronic Sphygmomanometer
CONTEC08A,Pulse Oximeter probe ESC0029 needs the help of the FST0002 Pulse Oximeter probe extension cable to be connected in
to the jack of the Patient Monitor PM50. Pulse Oximeter probe ESC0029 needs the help of the FST0014 Pulse Oximeter probe
extension cable to be connected in to the jack of the Patient Monitor CMS8000(new model)
Pulse Oximeter Probe user manual
The special power is supplied by from the equipments of CONTEC Patient Monitor 、Pulse Oximeter
Model explanation
Reusable adult
finger-clip SpO2 probe
(ESA0008,ESA0004,
ESA0005,ESA0014,
ESA0061)
Reusable child finger-clip
SpO2 probe
(ESA0015,ESA0016)
Reusable wrap SpO2
probe(type Y probe)
(ESC0029)
Applied crowds
Weight>40Kg adult
Weight10~40kg
Weight3-10kg
Figure 1
placement
Recommendatory
placement:forefinger
Recommendatory
child
placement:forefinger
Recommendatory
neonate
placement:sole of foot
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ESA0008 and is the answer not in the manual?

Questions and answers

Summary of Contents for CONTEC MEDICAL SYSTEMS ESA0008

  • Page 1 CONTEC MEDICAL SYSTEMS CO.,LTD Pulse Oximeter Probe user manual Product name: Pulse Oximeter Probe Product model: ESA0008,ESA0015,ESC0029,ESA0004,ESA0014,ESA0005,ESA0016,ESA0061 Scope of application: This product is used to match CONTEC Patient Monitor 、Pulse Oximeter and Electronic sphygmomanometer ,collect and transmit the signal from patient with continuance and no trauma. It is inapplicable to monitor the weak perfusion moving state and monitor for long, so check the measuring position or change for another position per 4 hours.
  • Page 2: Notice Items

    CONTEC MEDICAL SYSTEMS CO.,LTD Notice items: 1) pulse oximeter probe placement, the position without ductus arteriosus, BP cuff and vein input pipe is top-priority. 2) If the pulse oximeter probe can't monitor the state of pulsation, it shows that the position of probe is improper, or the position is too thick, too thin or having too deep pigment to reach a proper translucidus effect.
  • Page 3 CONTEC MEDICAL SYSTEMS CO.,LTD Explanation about graphs and symbols used on the product: Follow Keep in a Date of instructions Keep away cool, dry manufacture for use from sunlight place TYPE BF Covering Product WEEE IPX1 APPLIED Protection code disposal...
  • Page 4: Guidance And Manufacturer's Declaration - Electromagnetic Emission

    CONTEC MEDICAL SYSTEMS CO.,LTD Guidance and manufacturer’s declaration – electromagnetic emissions-for pulse oximeter probe Guidance and manufacturer’s declaration – electromagnetic emission is intended for use in the electromagnetic environment specified below. The pulse oximeter probe customer of the user of the should assure that it is used in such and environment.
  • Page 5: Guidance And Manufacturer's Declaration - Electromagnetic Immunity

    CONTEC MEDICAL SYSTEMS CO.,LTD Guidance and manufacturer’s declaration – electromagnetic immunity –for pulse oximeter probe that are not LIFE-SUPPORTING Guidance and manufacturer’s declaration – electromagnetic immunity The pulse oximeter probe is intended for use in the electromagnetic environment specified below. The customer or the user of pulse oximeter probe should assure that it is used in such an environment.
  • Page 6 NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. CONTEC MEDICAL SYSTEMS CO., LTD Address: No.112 Qinhuang West Street, Economic&Technical Development Zone,...