Test Equipment; Test Conditions - National Instruments SCXI-1121 Calibration Procedure

Hide thumbs Also See for SCXI-1121:
Table of Contents

Advertisement

Test Equipment

Test Conditions

© National Instruments Corporation
NI recommends that you use the equipment in Table 1 to calibrate the
SCXI-1121. If these instruments are not available, use the requirements
listed in Table 1 to select a suitable substitute.
Equipment
Calibrator
DMM
Resistors
If you do not have custom connection hardware, you need the following
connectors:
Terminal block, such as the SCXI-1320
Shielded 68-pin connector cable
50-pin ribbon cable
50-pin breakout box
SCXI-1349 adapter
These components give easy access to the individual pins on the
SCXI-1121 front and rear connectors.
Follow these guidelines to optimize the connections and environment
during calibration:
Keep connections to the SCXI-1121 short. Long cables and wires can
act as antennas, picking up extra noise and thermal offsets that affect
measurements.
Use shielded copper wire for all cable connections to the SCXI-1121.
Use twisted-pair wire to eliminate noise and thermal offsets.
Maintain the ambient temperature between 18 °C and 28 °C.
Keep relative humidity below 80%.
Allow a warm-up time of at least 15 minutes for the SCXI-1121 to
ensure that the measurement circuitry is at a stable operating
temperature.
Table 1. Test Equipment
Recommended
Model
Fluke 5700A
NI 4070
3
Requirements
50 ppm
5 1/2-digit, 15 ppm
120 Ω and 800 Ω, ±10%
SCXI-1121 Calibration Procedure