Configuring the Excitation Jumpers

You can configure each excitation channel of the SCXI-1121 to either a
voltage or current excitation mode. Each channel has two jumpers for this
purpose. Set both jumpers in the same mode for correct operation of the
excitation channel. Refer to Table 6 to determine how to set up the
SCXI-1121 in the desired mode. The SCXI-1121 ships to you in the voltage
mode.
Table 6. Voltage and Current Mode Excitation Jumper Settings
Excitation
Channel
Jumpers
0
W14 and
W15
1
W22 and
W23
2
W34 and
W35
3
W46 and
W47
Table 5. Filter Jumper Settings
4 Hz
Setting)
4 kHz
W5-A
W5-B
W21-A
W21-B
W31-A
W31-B
W43-A
W43-B
Voltage Mode
(Factory Setting)
W14
1
2
3
1 2
W22
1 2 3
1 2 3
W34
1 2 3
1 2 3
W46
1 2 3
1 2 3
9
Second Filter Jumper
4 Hz
(Factory
Setting)
4 kHz
W6
W8
W10
W12
Current Mode
W15
W14
1 2 3
1 2 3
3
W22
W23
1 2 3
1 2 3
W35
W34
1 2 3
1 2 3
W47
W46
1 2 3
1 2 3
SCXI-1121 Calibration Procedure
W7
W9
W11
W13
W15
W23
W35
W47