National Instruments SCXI-1121 Calibration Procedure page 18

Hide thumbs Also See for SCXI-1121:
Table of Contents

Advertisement

Gain
2000
2000
* Value available only when used with the SCXI-1327 high-voltage terminal block
SCXI-1121 Calibration Procedure
Table 12. SCXI-1121 Test Limits (Continued)
Test Point (V)
Limit (V)
0.00000
0.022238
–0.00225
–4.448611
4 Hz Filter Setting
Upper
Lower
Limit (V)
–0.022238
–4.551389
18
10 kHz Filter Setting
Upper
Lower
Limit (V)
Limit (V)
0.029480
–0.029480
–4.441369
–4.558631
ni.com

Advertisement

Table of Contents
loading

Table of Contents