Required Software
Calibrating the PXIe-5164 requires you to install the following software on the calibration
system:
•
NI-SCOPE 16.1 or later. PXIe-5164 was first supported in NI-SCOPE 16.1.
•
NI LabVIEW Instrument Design Libraries for Reconfigurable Oscilloscopes 16.1 or later.
PXIe-5164 was first supported in NI LabVIEW Instrument Design Libraries for
Reconfigurable Oscilloscopes 16.1.
Note
Instrument Design Libraries for Reconfigurable Oscilloscopes 16.2 or later.
•
Supported application development environment (ADE)—LabVIEW
You can download all required software from ni.com/downloads.
Related Documentation
For additional information, refer to the following documents as you perform the calibration
procedure:
•
PXIe-5164 Getting Started Guide
•
NI High-Speed Digitizers Help
•
NI Reconfigurable Oscilloscopes Help
•
PXIe-5164 Specifications
Visit
ni.com/manuals
Test Equipment
Refer to the following table for a list of necessary equipment and model recommendations for
calibration of the PXIe-5164.
If you do not have the recommended equipment, select a substitute calibration standard using
the specifications listed in the minimum requirements column of the table.
2 | ni.com | PXIe-5164 Calibration Procedure
To perform adjustment procedures, you must use NI LabVIEW
for the latest versions of these documents.