Configuring The Gain Jumpers - National Instruments SCXI-1121 Calibration Procedure

Hide thumbs Also See for SCXI-1121:
Table of Contents

Advertisement

© National Instruments Corporation
Table 2. Completion Network Jumpers
Channel
Jumpers
0
W1 and W2
1
W17 and
W18
2
W27 and
W28
3
W39 and
W40

Configuring the Gain Jumpers

Each input channel has two user-configurable gain stages. The first-stage
gain provides gains of 1, 10, 50, and 100. The second-stage gain provides
gains of 1, 2, 5, 10, and 20. The SCXI-1121 ships with the first-stage gain
set to 100 (position A) and the second-stage gain set to 10 (position D).
To change the gain setting of a specified channel on the SCXI-1121, refer
to Table 3 for the gain jumper reference designators. Refer to Figure 3 for
the location of the gain jumpers. Move the jumper to the position indicated
in Table 4.
Enable
Completion
A
B
1 2 3
W18
A
B
1 2 3
W28
A
B
1 2 3
W40
A
B
W39
1 2 3
7
Disable Network
(Factory Setting)
W2
A
B
W1
1 2 3
A
B
W17
1 2 3
l
A
B
W27
1 2 3
W40
A
B
1 2 3
SCXI-1121 Calibration Procedure