○Cross reference for troubleshooting
Reference
Problem
No.
There is no
1
display.
The display
appears, but
2
immediately
disappears.
An error display
3
is shown.
The displayed
4
value does not
change.
The displayed
5
value is
unstable.
The zero point
6
is displaced.
The back light
7
does not turn
on.
Possible cause
The battery is inserted
Check the orientation of
incorrectly.
battery.
The battery is running
Check how much battery is
out.
left.
The battery is running
Check for battery low display. Replace with a new battery.
out.
The auto power-off
Check the operating time has
mode is working.
passed 5 minutes.
A charge over ±5%F.S.
of the default zero
The display during zero clear
point is given to the
operation is ±5%F.S. at max.
sensor during zero
clear operation. (Er1)
The sensor is broken.
(Er2)
Error in internal data.
(Er3)
The upper/lower limit
Check the measured value is
of the measurable
displayed when the charged
voltage range is
object is kept far away from
exceeded.
the sensor.
(HHH/LLL)
The measuring
Check the measuring
distance is not
distance referring to the label
appropriate.
on the sensor (50 mm).
(HHH/LLL)
Check whether the LCD is
The peak hold mode
displaying "P" for peak hold
or bottom hold mode is
mode or "b" for bottom hold
set.
mode.
The ground wire is not
Check the ground wire is
grounded.
connected properly.
The ground wire is not
Check the ground wire is
grounded.
connected properly.
Measure in an open place
Ambient environment
like outside of a room, and
is charged.
check the zero point returns.
Operating
Check the back light is on in
environment is too
a dark place.
bright to see.
Investigation method
-15-
Countermeasure
Check the orientation of battery as
shown in the case and if it is
incorrect, reinsert the battery.
Replace with a new battery.
Press the POWER button with the
power supply off for 6 s. at least, and
extend the continuous operation
time to 15 minutes.
Return to a condition with no
charge, and retry zero clear
operation.
Stop using immediately and contact
SMC.
Turn off the power supply once, and
turn it on again.
Ask SMC to investigate if the error
persists.
If the measured value is displayed
normally for the object far away fro
the sensor, it means the charge
amount of the measured target is
beyond the specifications of the
product. Stop using immediately.
Adjust the distance between the
measured target and sensor to the
rated value (50 mm).
Hold the POWER button pressed
until "P" or "b" disappears, and then
move to the instantaneous display
mode.
Ground the ground wire.
Ground the ground wire.
Execute zero clear operation.
If the light is confirmed to be off, ask
SMC to investigate.
No.SFOD-OML0002-A