OBL PRO Series Use And Maintenance Manual page 29

Table of Contents

Advertisement

EFL09
FIRMWARE
RTC10
ELECTRONICS
HARDWARE
LOG11
FIRMWARE
LVL12
PROCESS
OVR13
PROCESS
RMT14
PROCESS
AMA15
PROCESS
AVA16
PROCESS
UVG17
ELECTRONICS
UVB18
HARDWARE
3.6 Analog Mode Calibration Procedure
The scope of the following procedure is to calibrate analog output and input mA signal to let the
pump recognize 4 mA and 20 mA. When the measured value by PRO KIT is quite different from
the input value (at least 0,3 mA difference), this procedure will help the accuracy of 4-20 mA
signal.
Disable Analog Output
The first operation to be accomplished is to disable analog mA output:
1) Go to the menu "settings→Analog output"
2) Set the analog output to "NO" clicking on it
File: UT-5601
Copyright © - OBL Metering pumps - All rights reserved
Log write error and FW
EXT FLASH
update error
RTC
Clock
power-on (the clock is
set to a non-admitted
time). Flat battery.
LOG
Error in recording the
file log
LEVEL
STOP input of the level
sensor
OVERFLOW
Overflow alarm
REMOTE STOP
Alarm resulting from
remote stop
ANALOG mA
Input current beyond
the range set in the
analog mA mode.
ANALOG V
Input voltage beyond
the range set in the
analog V mode.
Power supply voltage
too low.
YELLOW
error
upon
FIXED RED
FIXED RED OR
FIXED YELLOW
FIXED RED
FIXED RED
FIXED RED
28