The Transfer Extension and the Radiolucent Extension
The transfer extension is used to extend the table when putting a patient on or taking a patient
off the tabletop. It is not used for imaging. The radiolucent extension, significantly longer than
the transfer extension, has a 31" radiolucent C-Arm area with no metal cross members. Both
extensions are installed using the same procedure. Once the Radiolucent Extension is in use,
the patient load rating is limited to 400 lb (181.4 kg).
To Install the Extension:
Open the vertical doors at the foot end of the table (the doors pivot toward the outside of
the table.)
Pick up the extension from the end and slide the silver, tapered ends into the openings
created by opening the vertical doors.
Slide the extension fully into the table until you feel the extension lock into place.
To Remove the Extension:
NOTE: The x-ray grid, film tray and drain pan can remain in place if using the Transfer
Extension but must be removed if the Radiolucent Extension is used.
Stand at the foot end of the table and press the extension release levers located under the
middle of each leg extension rail. The extension can be slid out from the table.
Close the vertical doors.
Figure 2.4. Insert Either Table Extension into the Extension Receiving Tubes. To

Disposable Skirt (Optional)

The Biodex Urology & Brachytherapy C-Arm Tables (models 058-800 and 058-810) can be
ordered with the optional Plastic Sanitary Skirts to protect the table bellows from patient
blood and other liquids. The table has plastic clips located immediately above the bellows.
Secure the skirt to the clips as described below:
1. Slide the plastic skirt underneath the clip at the foot end of the table and lower the clip
pressing along the entire length of the clip to hold the skirt in place.
2. Work around the table, tucking the top edge of the plastic skirt under each clip and
lowering the clip to secure the plastic until the bellows are completely covered. Be sure
to overlap the skirt across the entire length of the table's foot end.
Vertical Door
Remove the extension from the Table, Press the Extension Release Levers
and Slide the Extension Out of the Table.