Preparing For Calibration; Accessing The Calibration/Test Menu - GE DigitalFlo XGM868i Service Manual

Panametrics general-purpose gas flow transmitter (1 & 2 channel)
Table of Contents

Advertisement

1.4.1 Preparing for Calibration

Prepare for the calibration procedure by connecting a Slot 0 analog output (or an independent calibrated current source)
to the desired input on the option card. See Figure 4 on page 12 to identify the IN(+) and RTN(-) pins for the desired
input on terminal block J2.
Note: The zero point of the analog input may be set for either 0 mA or 4 mA. However, the calibration always uses the
4 mA point, as the meter extrapolates this value to obtain the 0 mA point.

1.4.2 Accessing the Calibration/Test Menu

1. From the Keypad Program, press the right arrow key to scroll to the CALIB menu. Press [Enter].
2. In the Cal window, scroll to Slot 1. Press [Enter].
3. Scroll to the desired Input to open the ANALOG INPUT menu. (
Press [Enter].
Calibrating
IMPORTANT:
source on terminal block
4. Scroll to the appropriate option to select the reference point to be calibrated. Press [Enter].
The procedure for calibrating all the inputs is the same. However, when calibrating a different input, be
IMPORTANT:
sure to reconnect the current source at terminal block J2. See Figure 4 on page 12 for the correct pin
numbers.
5. Do one of the following:
Proceed to<Title Font> 4 mA Option on page 5.
Proceed to<Title Font> 20 mA Option on page 5.
1.4.2a
4 mA Option
Set the calibrated current source to 4 mA.
If you selected the 4 mA option in the ANALOG INPUT menu, scroll to STORE to accept the current 4 mA value or
on ABORT to cancel the entry. Press [Enter]. In either case, the XGM868i returns to the Analog Input window.
1.4.2b
20 mA Option
Set the calibrated current source to 20 mA.
If you selected the 20 mA option in the ANALOG INPUT option, scroll STORE to accept the current 20 mA value
or to ABORT to cancel the entry. Press [Enter]. In either case, the XGM868i returns to the ANALOG INPUT
window.
Press [Escape] to return to the Slot 1 window and calibrate additional inputs/outputs, or press [Escape] again to
close the CALIB menu.
DigitalFlow™ XGM868i Service Manual
or
is identical to calibrating
Input B, C
D
. Refer to Figure 4 on page 12 for the correct pin numbers.
J2
will be used as an example in this manual.)
Input A
. However, be sure to reconnect the current
Input A
Chapter 1. Calibration
5