Chapter 1. Calibration
1.3.4 Testing Output Linearity (cont.)
Table 1 lists the expected ammeter readings at various % Full Scale settings, for both 4-20 mA and 0-20 mA scales.
Refer to this table to verify the accuracy of the ammeter readings taken above.
If the linearity test readings are not within 5 A of the values listed in Table 1, check the accuracy and wiring of the
ammeter. Then, repeat the low and high end calibrations. If the analog output still does not pass the linearity test,
contact the factory for assistance.
10. Scroll to EXIT and press [Enter] to return to the Cal menu. Then choose another output to calibrate or press
[Escape] to leave the CALIB menu.
This completes calibration of the
inputs/outputs.

1.4 Calibrating Analog Inputs

Analog inputs may be added to the Model XGM868i by installing an appropriate option card in Slot 1. This option card
contains two or four analog inputs, which are designated as A, B, C and D. Both the zero-point and full-scale values for
each input must be calibrated.
Calibration of the analog inputs requires the use of a calibrated current source. If an independent calibrated current
source is not available, one of the Slot 0 analog outputs may be used for the calibration. During the analog input
calibration, the Slot 0 analog output will supply the low reference, high reference, 4 mA and
20 mA signals at the appropriate times.
If a Slot 0 analog output will be used to calibrate the analog inputs, the Slot 0 analog output calibration
IMPORTANT:
procedure must be completed first.
4
Table 1: Expected Ammeter Readings
% Full Scale
4-20 mA Scale*
0
10
20
30
40
50
60
70
80
90
100
* all ammeter readings should be ±0.005 mA
analog outputs. Proceed to the appropriate section to calibrate additional
Slot 0
0-20 mA Scale*
4.000
5.600
7.200
8.800
10.400
12.000
10.000
13.600
12.000
15.200
14.000
16.800
16.000
18.400
18.000
20.000
20.000
0.000
2.000
4.000
6.000
8.000
DigitalFlow™ XGM868i Service Manual