TECHNICAL SPECIFICATIONS
Electrical Characteristics
Power Input
Voltage: 100–230 VAC, 50/60 Hz
Power: 1.0 Amp max
Fuse Rating
1 Amp x 250 volts Slo-Blo (Time Delayed) fuse
Display
5 volts dc, max 0.5 amp, 8 lines x 20 characters per line presentation, LCD with white LED
back light.
Keyboard
Membrane keypad with 20 keys
USB Flash Drive Interface
USB 2.0 (low power) Flash Drive interface to exchange Programs
Physical Characteristics
Dimensions:
Width:
Depth:
Height:
Weight:
Actual Weight:
Shipping Weight: 110 lbs (52 kg)
Operating Environmental Conditions
All Compass Stainer products are laboratory systems and should be used ONLY in laboratories with
the following conditions:
Indoor use Only
Ambient Temperature 15
Relative Humidity
Ambient Pressure
Altitude
page 6
32" (81 cm)
Additional required clearance 6" (15 cm)
31" (79 cm)
Additional required clearance 6" (15 cm)
16" (41 cm)
Additional required clearance 12" (30 cm)
55 lbs (25 kg) Max – without reagents in the container
o
o
o
C to 35
C (59
F to 95
20-80% non-condensing
29"-31" (740-790 mm) of Hg
Less than 6500 feet (Less than 2000 Meters)