Staining Protocol, 95% Reagent Alcohol Fixed ThinPrep
for use with ThinPrep Imaging
Equipment: Hologic Compass Stainer
Run no more than four slide racks together. Wait for all four to be unloaded before adding new racks.
Caution: Change the Distilled / Instrument Feed Water bath at Station 5 after every four racks of slides.
Maintain bath heights to completely cover the slides at full immersion.
Note:
Bath fill volume = 360 ml.
Station
Time
LOAD
n/a
1
01'00"
2
01'00"
3
01'00"
4
07'00"
5
00'10"
6
01'00"
7
00'30"
8
00'30"
9
00'30"
10
00'30"
11
00'30"
12
02'00"
13
00'15"
14
00'15"
15
04'00"
16
01'00"
17
01'00"
18
00'30"
19
00'30"
20
00'30"
21
01'00"
22
03'00"
UNLOAD
n/a
a. Hologic specification for IFW: ≥1.0 megohm-cm Resistivity OR ≤1.0 μSiemens/cm Conductivity (Refer to Clinical and Laboratory Standards
Institute (CLSI) document C3-A4, 2006)
b. See ThinPrep Stain User's Manual, Section 1, "STAINING" or contact Hologic for current list of Hologic approved clearing agents and mounting
medias
Agitation: Dips = 10, Height = 30 mm
Duration
n/a
95% Reagent or Ethyl alcohol
Non-critical
70% Reagent or Ethyl alcohol
Non-critical
50% Reagent or Ethyl alcohol
Non-critical
Distilled or Instrument Feed Water (IFW)
Critical
ThinPrep Nuclear Stain
Critical
Distilled or Instrument Feed Water (IFW)
racks.)
Critical
ThinPrep Rinse Solution
Non-critical
Distilled or Instrument Feed Water (IFW)
Critical
ThinPrep Bluing Solution
Non-critical
Distilled or Instrument Feed Water (IFW)
Non-critical
50% Reagent or Ethyl alcohol
Non-critical
95% Reagent or Ethyl alcohol
Critical
ThinPrep Orange G Solution
Critical
95% Reagent or Ethyl alcohol
Critical
95% Reagent or Ethyl alcohol
Critical
ThinPrep EA Solution
Critical
95% Reagent or Ethyl alcohol
Critical
95% Reagent or Ethyl alcohol
Non-critical
100% Reagent or Ethyl alcohol
Non-critical
100% Reagent or Ethyl alcohol
Non-critical
100% Reagent or Ethyl alcohol
Non-critical
Xylene or other Hologic approved clearing agent