TROUBLESHOOTING GUIDE
Symptom
The unit does not
power up
Display does not power
up
Up-down mechanism
does not initialize or
move
Rotary mechanism does
not initialize or move
page 44
Probable Causes
Unit not connected to the
power outlet
No power at the power
outlet
Power Switch in the OFF
position
Blown Fuse
Communication between
host and the Display
controller not working or
Display power is
interrupted
Obstruction in the transport
mechanism, mechanism
does not move freely
Defective PCB
Defective Motor
Broken Belt
Follow the same procedure
as described above for up-
down mechanism
Compass
Stainer Operator's Manual
Suggested Actions
Ensure that the power cord is
plugged in at the outlet as well as in
the rear of the unit.
Check the circuit breaker for the
power outlet and ensure that the
power is available at the power
outlet.
Turn ON the power switch in the
rear of the.
Check the fuses and replace any
blown fuse with a fuse of the same
rating. Investigate the cause for
overload and correct the problem
before re-starting the system.
Contact Hologic Technical Support
Remove any obstructions
Contact Hologic Technical Support
Contact Hologic Technical Support
Contact Hologic Technical Support
Contact Hologic Technical Support
Follow the same procedure as
described above for up-down
mechanism