TROUBLESHOOTING GUIDE
Symptom
The unit does not
power up
Display does not power
up
Up-down mechanism
does not initialize or
move
Rotary mechanism does
not initialize or move
page 44
Probable Causes
•
Unit not connected to the
power outlet
•
No power at the power
outlet
•
Power Switch in the OFF
position
•
Blown Fuse
•
Communication between
host and the Display
controller not working or
Display power is
interrupted
•
Obstruction in the transport
mechanism, mechanism
does not move freely
•
Defective PCB
•
Defective Motor
•
Broken Belt
•
Follow the same procedure
as described above for up-
down mechanism
™
Compass
Stainer Operator's Manual
Suggested Actions
•
Ensure that the power cord is
plugged in at the outlet as well as in
the rear of the unit.
•
Check the circuit breaker for the
power outlet and ensure that the
power is available at the power
outlet.
•
Turn ON the power switch in the
rear of the.
•
Check the fuses and replace any
blown fuse with a fuse of the same
rating. Investigate the cause for
overload and correct the problem
before re-starting the system.
•
Contact Hologic Technical Support
•
Remove any obstructions
•
Contact Hologic Technical Support
•
Contact Hologic Technical Support
•
Contact Hologic Technical Support
•
Contact Hologic Technical Support
•
Follow the same procedure as
described above for up-down
mechanism