• Always inspect support surface covers (top and bottom) for tears, punctures, excessive wear, and misaligned zippers
before each use. If compromised immediately remove the support surface from service.
• Always make sure that you wipe each product with clean water and thoroughly dry each product after any application of
chemical solutions. Some chemical solutions are corrosive in nature and may cause damage to the product if you use
them improperly. If you do not properly rinse and dry the product, a corrosive residue may be left on the surface of the
product that may cause premature degradation of critical components. Failure to follow these instructions may void your
warranty.
• Always apply the supplied Velcro to the 0747 stretcher to make sure that the mattress is secure. Non-use may result in
patient harm due to mattress movement.
C C A A U U T T I I O O N N
• Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this
manual.
• Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids its warranty.
• Always be aware of devices or equipment that are placed on the top of the support surface. Damage to the surface may
occur due to the weight of the equipment, heat generated by the equipment, or sharp edges on the equipment.
• Do not put overlays or accessories inside the cover to avoid the risk of reducing pressure redistribution performance.
• Always evaluate the appropriate Cardiopulmonary Resuscitation (CPR) protocol to be used with this product before
operating.
• Do not allow liquid to seep into the zipper area or watershed cover barrier when washing the underside of the support
surface. Fluids allowed to come in contact with the zipper may leak into the support surface.
• Do not iron, dry-clean, or tumble dry the support surface covers.
• Do not power wash the support surface as this may damage the product.
• Always completely dry the support surface covers before storing, adding linens, or placing a patient on the surface.
Drying the product helps to prevent the performance of the product from being impaired.
• Do not over expose the covers to higher concentration chemical solutions as these may degrade the covers.
• Do not use accelerated hydrogen peroxides or quaternaries that contain glycol ethers as they may damage the cover
and reduce the legibility of the graphics.
• Failure to follow manufacturing instructions may also affect useful life of the support surface cover.