Manual Disinfection - Pari 041G0500 Instructions For Use Manual

Filter/ventil set
Table of Contents

Advertisement

Available languages

Available languages

PROCEDURE
• Place all the parts in the cleaning solution.
• Clean the parts thoroughly with a brush.
• Thoroughly rinse the parts under hygienically tested water
and leave them to dry completely on a clean, dry and
absorbent surface.
VALIDATION
The method has been validated using:
®
– Korsolex
Endo Cleaner 0.5% (Bode)
– approx. 50 °C warm water
– Duration of cleaning procedure: approx. 5 minutes

Manual disinfection

For manual disinfection, select a suitable instrument disinfectant
that has been proven to be effective and compatible with the
cleaning agent used.
EQUIPMENT
– Disinfectant
– Water
PARI product components are not ...
CAUTION
PARI product components are not suitable for treatment with
instrument disinfectants based on quaternary ammonium
compounds. Quaternary ammonium compounds tend to
become concentrated in plastic materials, which can cause
incompatibility reactions in patients.