Error Codes
12.
Error Codes
(— LED Error lights for a long time, • LED Error lights briefly)
Flashing
sequence
..
Mode not found.
2x short
..-
White error flash
2x short, 1x long
...
Flash full
3x short
....
Loading suppressed due to active
data transmission.
4x short
--
Data overflow in the sensor
2x long
--.
Data overflow during receipt of the
data from the sensor.
2x long, 1x short
--..
Data overflow for serial port RS422
2x long, 2x short
scanCONTROL 30xx
Cause
Group: Loading / saving configuration
Select different one.
Contact manufacturer, return sensor.
None, contact manufacturer.
Stop active data transmission.
Group: Data processing and transmission
Select smaller measuring field, reduce
profile frequency, select less complex
measuring program.
Select smaller measuring field, reduce
profile frequency, select less complex
measuring program.
Reduce profile frequency, select less
complex measuring program.
Remedy
Only previously stored modes
can be called up.
Should not occur in normal
operation.
Should not occur in normal
operation.
Prevents PC software crashes.
Data can be impaired; exposure
time can be longer than ex-
pected.
Data can be faulty.
Data can be faulty.
Notes
Page 57