When averaging mode is tuned on, period for signal updating becomes longer.
а) averaging mode is turned on
Figure 2.63 – Measurement mode with signal averaging - «AVG» parameter
2.4.11.3 «Auto ZOOM» mode
«Auto ZOOM» mode allows to focus scanning time on the signal, that crosses the gate. To turn on
the mode user has to select «ON» in «Auto ZOOM» parameter, in «SETTINGS».
Focusing on A-gate is made by touching the symbol of gate signalization «
focusing on B-gate - by touching the symbol «
а) selection of «Auto ZOOM» mode for A-gate
To turn off «Аuto ZOOM» mode user has to touch once more the symbol of the gate signalization.

2.4.12 Operating with device memory

There are two modes for operating with device memory: «ARCHIVE» and «RESULTS».
2.4.12.1 Saving of measurement results and device settings
Saving of the current settings or measurement results is made in «FLAW DETECTOR» mode by
pushing key «
» or «SAVE». After this, panel for saving type selection is appeared on the screen
(Figure 2.65).
User selects the required type of saving by navigation keys « », « » and confirm the selection
by pushing key «
» or touching the selected type by finger on the screen. After type selection of
saving keyboard appears on the screen for entering the name of saving (Figure 2.66).
Ultrasonic Flaw Detector NOVOTEST UD3701
».
Figure 2.64 – «Auto ZOOM» mode for A-gate
b) averaging mode over 8 signals
b) «Auto ZOOM» mode for A-gate
Operation manual
» (Figure 2.64, а),
Page 52