3.15.2 Anti-tipper swing-away

Height adjustment / Folding:
A
B
C
1. To adjust the height, press on the two knobs B placed
on each side of the housing.
2. Adjust the anti-tipper tube C to the desired height.
Make sure that the anti-tipper is fitted correctly,
the two knobs B must lock into position.
3. To fold the anti-tipper, press the top of the housing A
downwards.
4. Push the anti-tipper sideways.
Do not use the knobs B to fold or unfold the
anti-tipper.
CAUTION!
Risk of pinching fingers
– Use the foot to fold the anti-tipper,
otherwise there is a risk of pinching
the fingers between the housing of the
anti-tipper and the tube of the anti-tipper.
5. To activate the anti-tipper, push the anti-tipper back to
its original place, it will automatically lock into position.
Depth adjustment:
1. Press on the knob D.
2. Adjust the inner wheel tube to the desired position.
Make sure that the tube is fitted correctly on
each side (if applicable), the knob D must lock
into position.
This adjustment is required with reference to the
position and diameter of the rear wheel as well as
the user conditions and his particular safety limits.
WARNING!
Risk of overturning
– The adjustments of the anti-tipper must be
carried out by a qualified technician.
16

3.15.3 Anti-tip device ultra low

Mounting / Activating the anti-tip device
1. Remove the plastic plugs from the end tube of the
chassis A.
2. Push the button B to insert the anti-tip device into the
end tube A.
3. Make sure that the button B is securely locked in its
position.
The ultra low anti-tip device are used for seat height
360 mm.
Deactivating the anti-tip device
1. Push the button B.
2. Turn the anti-tip device C 180°.
The wheels of the anti-tip device should be
pointing upwards.
3. Make sure that the button B is securely locked in its
position.

3.16 Seat cushion

A suitable cushion is recommended to provide an even
pressure distribution on the seat.
CAUTION!
Risk of instability
Adding a cushion to the seat will raise your height
above the ground and can affect your stability in
all directions. If a cushion is changed it may also
change the user's stability.
– If the thickness of the cushion is changed, a
complete set up of the wheelchair need to be
done by a qualified technician.
– We recommend to use an Invacare or Matrx
cushion with anti-slip base cover in order to
prevent sliding.
A
B
1536533-F