Maintenance; Safety Information; Maintenance Schedule - Invacare Rea Spirea 4 NG User Manual

Medium active wheelchair
Table of Contents

Advertisement

Rea® Spirea® 4 NG

8 Maintenance

8.1 Safety information

WARNING!
Some materials deteriorate naturally over time.
This could result in damage to wheelchair
components.
– Your wheelchair should be checked by a
qualified technician at least once a year or if it
has not been used for a long period.
Packaging for return
The wheelchair shall be sent to the authorized
provider in an appropriate packaging to avoid damage
during the shipment.

8.2 Maintenance Schedule

To ensure safe and reliable operation, carry out the following
visual checks and maintenance regularly or have it carried
out by another person.
weekly
Check the tyre
pressure
Check that rear
wheels are seated
correctly
Check posture belt
Check the folding
mechanism
Check the castors
Check bolts
Check spokes
Check parking brakes
Have wheelchair
checked by a qualified
technician
General visual check
1. Examine your wheelchair for loose parts, cracks or other
defects.
2. If you find anything, stop using your wheelchair
immediately and contact an authorized provider.
Check the tyre pressure
1. Check the tyre pressure, see section "Tyres" for more
information.
2. Inflate the tyres to the required pressure.
3. Check the tyre tread at the same time.
4. If necessary, change the tyres.
Check that rear wheels are seated correctly
1. Pull on the rear wheel to check that the removable axle
is seated correctly. The wheel should not come off.
2. If the rear wheels are not engaged properly, remove
any dirt or deposits. If the problem persists, have the
removable axles re-fitted by a qualified technician.
36
monthly
annually
x
x
x
x
x
x
x
x
x
Check posture belt
1. Check that the posture belt is adjusted correctly.
IMPORTANT!
– Loose posture belts must be adjusted by an
authorized provider.
– Damaged posture belts must be replaced by
a qualified technician.
Check the folding mechanism
1. Check that the folding mechanism is easy operating.
Check the castors
1. Check that the castors turn freely.
2. Remove any dirt or hair from the castor bearings.
3. Faulty or worn out castors must be replaced by a
qualified technician.
Check the bolts
Bolts can work loose through constant use.
1. Check that the bolts are tight securely and without free
play (on the footrest, castor, castor housing, seat cover,
sides, backrest, frame, seat module).
2. Tighten any loose bolts with the suitable torque.
Therefore refer to the Service Manual, available on the
internet at www.invacare.eu.com.
IMPORTANT!
Self-locking screws/nuts or thread-locking
adhesive are used for several connections. If
these are loosened, they must be replaced by
new self-locking screws/nuts or they must be
secured using new thread-locking adhesive.
– Self-locking screws/nuts must be replaced by
a qualified technician.
Check the spoke tension
The spokes should not be loose or distorted.
1. Loose spokes must be tightened by a qualified technician.
2. Broken spokes must be replaced by a qualified
technician.
Check the parking brakes
1. Check that the parking brakes are positioned correctly.
The brake is set correctly if the brake shoe depresses the
tyre by a few millimeters when the brake is engaged.
2. If you find that the setting is not correct, have the
brakes correctly adjusted by a qualified technician.
IMPORTANT!
The parking brakes must be reset after replacing
the rear wheels or changing their position.
Checking after a heavy collision or blow
IMPORTANT!
The wheelchair can sustain visibly undetectable
damage as a result of a heavy collision or hard
blow.
– It is essential to have your wheelchair checked
by a qualified technician after a heavy collision
or hard blow.
1536533-F