Supplemental Oxygen; Setting Pressures And Features - DeVilbiss DV63 STANDARD PLUS Service Manual

Hide thumbs Also See for DV63 STANDARD PLUS:
Table of Contents

Advertisement

DV6x CPAP devices will operate on 12V DC power that has been
inverted to 120V AC power. NOTE: See DC Power Chart below for
the correct inverter specifications.
DC Power Chart for Vehicle & 12V Marine Battery
Inverter
Configuration of
Specification for
Use
use in vehicle
If using a vehicle battery for multiple nights without powering the vehicle's
motor during use, power the vehicle during the day to replenish the battery.
CPAP device
120V AC Modified
with heated
or Quasi Sine
humidification
Wave Inverter with
setting of 5
200w capacity
120V AC Modified
or Quasi Sine
Wave Inverter with
300w capacity
230V AC Modified
or Quasi Sine
Wave Inverter with
400w capacity
CPAP device
without heated
humidification
E. SUPPLEMENTAL OXYGEN
WARNING-OXYGEN USE
Oxygen supports combustion. To avoid possible physical injury,
do not smoke while using this device with supplemental oxygen.
Do not use this device near hot objects, volatile substances or
sources of open flames.
Always start the CPAP blower before starting the oxygen flow.
Stop the oxygen flow before stopping the CPAP blower.
Never allow the oxygen source to run continuously while
connected to the DeVilbiss CPAP device if the device is not in
use. If the DeVilbiss CPAP device is not being used, turn off the
oxygen flow.
CPAP pressure must be set higher than 8 cmH
supplemental oxygen use to prevent patient breathing pattern
from forcing oxygen back into the CPAP.
At a fixed flow rate of supplemental oxygen, the inhaled oxygen
concentration will vary depending on the pressure setting, patient
breathing pattern, mask selection and leak rate.
This warning applies to most types of CPAP devices.
CAUTION
Oxygen is a prescribed gas and should only be administered
under the supervision of a physician.
The flow setting of the oxygen source must be specified by a
physician.
Maximum oxygen pressure is 50 psi (345 kPa). Maximum oxygen
flow rate is 10 LPM.
If a physician prescribed supplemental oxygen, it can be added one of
two ways:
Attach oxygen tubing directly to the oxygen port on the mask
Insert the optional oxygen adapter into the device's air supply
G E N E R A L I N F O R M AT I O N
Battery
Battery Hours
Specification for
@ 20 cmH
O
2
stand-alone use
120 amp-hour,
deep-cycle
8 hours
Marine Type
240 amp-hour,
16 hours
deep-cycle
Marine Type
80 amp-hour,
8 hours
deep-cycle
Marine Type
O for
2
port. Then, insert the air supply tubing and the oxygen tubing to
the adapter.
Always START CPAP before starting oxygen flow.
Always STOP oxygen flow before stopping CPAP.
Oxygen Adapter
(7353D-601)
F. SETTING PRESSURES AND
FEATURES
Factory Default Settings
Factory default settings are installed on the IntelliPAP 2/DeVilbiss
BLUE DV6x CPAP device prior to shipping.
Setting
Operating Mode
CPAP pressure
Upper Pressure Limit (UPL)
Lower Pressure Limit (LPL)
Apnea %
Apnea Duration
Hypopnea %
Hypopnea Duration
Ramp Pressure
Ramp Time
Split Night - Inhibit Time
Number of Sessions
Usage Threshold
Wakeup Alert
5
Default Value
DV63 - CPAP
DV64 - AutoAdjust
Default 10.0 (DV63 initial setting is 8.0,
default is 10.0)
15.0 for DV64 AutoAdjust mode
7.0 for DV64 AutoAdjust mode
90
10 seconds
40
10 seconds
4.0 cmH
O
2
20 minutes
240 minutes
0 (off)
4 hours
0 (Off)