2. Perform the calibration – select 100dB and use the slider to adjust the calibration.
3. If needed, perform the peSPL-nHL correction select 0dB and use the slider to adjust the correction.
a. Stimuli can also be saved and presented to use only SPL format
4. Make sure to do this operation for any transducer and both Right and Left side!