The latency and inter-latency calculations are indicated on the printout.
Please consult local requirement for maximum latency differences across ears.
Be aware to compensate for single sided high-frequency hearing loss, correcting the average degree of
recruitment.
-
4kHz loss < 50dB nHL: no correction.
-
Deduct 0.1 ms per 10dB above 50dB nHL.
Selters & Brackmann (1977), also described in NHSP UK.
Please refer to the Neuro Latency functions described under Record and Edit tabs later in this instruction for use
manual and the quick guide Neuro Latency Testing with Eclipse in the Additional Information.
Further information about Neuro Latency recordings can be obtained from Eclipse Additional Information
Manual.
D-0120578-C – 2019/11
Eclipse – Instructions for Use - US
Page 47