10.2.1 peSPL to nHL correction values

ECochG/ABR15/ABR30/AMLR/RATE STUDY/VEMP 0 dB 2-2-2 cycle
linear envelope
Hz
Insert phone
250
28.0
500
23.5
750
21.0
1000
21.5
1500
26.0
2000
28.5
3000
30.0
4000
32.5
6000
36.5
8000
41.0
ECochG/ABR15/ABR30/AMLR/RATE STUDY/VEMP 0 dB
Insert phone
Click
35.5
NB CE-Chirp® LS
ECochG/ABR15/ABR30/AMLR/RATE STUDY/VEMP 0 dB
Hz
Insert phone
500
25.5
1000
24.0
2000
30.5
4000
34.5
ECochG/ABR15/ABR30/AMLR/RATE STUDY/VEMP 0 dB
Insert phone
31.5
Only tone burst correction values change for ALR & MMN testing. For Click and CE-Chirps
D-0120578-C – 2019/11
Eclipse – Instructions for Use - US
Toneburst
Headphone
38.0
25.5
23.0
21.5
23.0
24.5
26.5
32.0
37.5
41.5
ISO 389-6:2007
Click
Headphone
30.0
Headphone
27.5
24.0
26.5
34.0
CE-Chirp® LS
Headphone
26.5
Bone
Hz
Insert phone
74.5
250
69.5
500
61.0
750
56.0
1000
51,5
1500
47.5
2000
46.0
3000
52.0
4000
60.0
6000
65.5
8000
Bone
Insert phone
51.5
Click
Bone
Hz
Insert phone
74.0
500
61.0
1000
50.0
2000
55.0
4000
Bone
Insert phone
51.0
Toneburst
ALR/MMN dB 25-50-25 ms
Headphone
17.5
27.0
9.5
13.5
6.0
9.0
5.5
7.5
9.5
7.5
11.5
9.0
13.0
11.5
15.0
12.0
16.0
16.0
15.5
15.5
ISO 389-1:2000, ISO 389-2:1994, ISO 389-3:1994
Click
ALR/MMN 0 dB
Headphone
35.5
30.0
NB CE-Chirp® LS
ALR/MMN 0 dB
Headphone
25.5
27.5
24.0
24.0
30.5
26.5
34.5
34.0
CE-Chirp® LS
ALR/MMN 0 dB
Headphone
31.5
26.5
®
LS, the same correction is applied.
Bone
67.0
58.0
48.5
42.5
36.5
31.0
30.0
35.5
40.0
40.0
Bone
51.5
Bone
74.0
61.0
50.0
55.0
Bone
51.0