Electrical Zero Adjustment - Ametek 120HD User Manual

Portable hydrogen analyzer
Table of Contents

Advertisement

Figure 3-1.
Front display with
adjustment knobs.
3-2 | Model 120HD Hydrogen Analyzer

Electrical zero Adjustment

This adjustment must be made before the analyzer is used for the first
time but is necessary only occasionally after that.
NOTE
For the initial electrical zero adjustment and all subsequent electrical zero adjust-
ments, if the digital display does not read zero, continue with the steps below.
1. Provide a low pressure (<5 PSIG), low flow rate source of air/nitrogen to
purge the gas passages in the analyzer of any residual gases from previous
tests.
2. Check and adjust the CURRENT ADJUST, if necessary.
3. Turn Knob 1 to the TEST position needed - either standard display or
backlit display (Figure 3-1).
4. Flow air or nitrogen through the analyzer.
5. The digital readout should settle at 0.0 within five minutes. If it does
not, adjust the Electrical Zero Adjust Screw #4 (Figure 3-1) until the
reading is 00.0. The 120HD is now ready to analyze a gas sample.
Displays
Sensor Current
(Backlight)
Electrical Zero
Adjustment Screw
CURRENT ADJUST
Current Adjustment
Knob
(FOR CALIBRATION ONLY)
Displays % H2
(No Backlight)
Displays % H2
1
(Backlight)
4
SPAN ADJUST

Advertisement

Table of Contents
loading

This manual is also suitable for:

00135086

Table of Contents