Step
to GD10P manual).
5c.
Position the magnet over
5d.
The displayed value keeps changing until it stabilizes. Wait
approximately 2 minutes for the stabilization of the measure.
5e.
As soon as the instrument stabilizes at a value, place the magnet
over
to exit the sensitivity adjustment function. Continue
to step 6a.
5.4.4 Validating your calibration
Step
6a.
The display indicates «CnF» (Confirmation).
6b.
Position the magnet on
6c.
The display indicates «nOn» (No).
6d.
To validate and confirm your calibration values, set the magnet
on
to change No into Yes and then on
confirm. Continue as under paragraph End of zero-point
adjustment and calibration.
6e
Otherwise place the magnet on
When you do so, the detector will return to normal mode after a
one-minute countdown and without applying any of the previous
adjustments.
NPO60GB
Revision F.2
Action
.
Action
.
.
OLCT 60